Instrument Emulations - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Stop Bits ← RS232 using USB adapter
Sets the number of stop bits for the serial remote control interface.
Remote command:
:SYSTem:COMMunicate:SERial:SBITs
Visa Resource Strings
Displays the visa resource strings, used for remote control of the instrument. Each
interface requires an individual unique address, to identify the instrument for remote
control.
Remote command:
:SYSTem:COMMunicate:HISLip:RESource?
:SYSTem:COMMunicate:NETWork:RESource?
:SYSTem:COMMunicate:SOCKet:RESource?
:SYSTem:COMMunicate:GPIB:RESource?
:SYSTem:COMMunicate:USB:RESource?
:SYSTem:COMMunicate:SERial:RESource?
Goto Local
Switches the instrument to operate in local control mode.
Switching from remote to local control mode can be also done with one of the following
actions:
manually with the LOCAL key on the front panel
with the interface command &GTL via the remote control interface
with the key combination CTRL + Q.
Remote command:
&GTL
5.2.3.15

Instrument Emulations

It is also possible to remotely control the R&S SMB via the command set of another
signal generator, as for example of an HP generator. With this function you can, for
example, replace a signal generator with an R&S SMB in an automated test setup,
without adjusting the command scripts used.
You find all the remote control command sets supported by the R&S SMB in a selec-
tion list. For more information on this topic, an application note describes in detail how
to use this feature. See the product site of the R&S SMB.
The selected instrument also defines the identification string that is retrieved with query
*IDN?. In addition to the preset values, you can enter a user-defined identification
string, for example to provide individual identification for each generator, like
'MY_R&S SMB' (see
As any other parameter, you can additionally change the remote control command set
to be emulated via the
emulation, the R&S SMB specific command set is disabled, that means this command
is no longer effective. To return, you need to know the corresponding remote control
command of the simulated instrument. If you emulate an HP generator for example,
the HP command EX returns to the SCPI command set.
Operating Manual 1407.0806.32 ─ 21
Mode
and
IDN
String).
Language
command. However, once you have switched to an
Instrument Function
General Instrument Settings
on page 443
on page 441
on page 441
on page 443
on page 441
on page 441
on page 442
110

Advertisement

Table of Contents
loading

Table of Contents