R&S SMB100A Operating Manual page 302

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Manual operation:
:CALibration<hw>:FREQuency[:MEASure]?
Starts all adjustments which affect the frequency.
Return values:
<Measure>
Example:
Usage:
Manual operation:
:CALibration<hw>:LEVel[:MEASure]? [<Force>]
Starts all adjustments which affect the level.
The acquired correction values improve the settling time and the signal quality.
Query parameters:
<Force>
Return values:
<Measure>
Example:
Usage:
Manual operation:
:CALibration<hw>:LEVel:EXTern:DATA <Data>
Queries what data has been used for the level calibration.
By default the instrument uses correction data obtained in the factory before delivery.
In addition, customer data can be used for external level correction. The customer data
is obtained using a R&S NRP power sensor. External level correction is a protected
function (see service manual, chapter 2, "Adjustment").
Parameters:
<Data>
Operating Manual 1407.0806.32 ─ 21
See
"Adjust FM Offset"
0 | 1
CAL:FREQ:MEAS?
starts the adjustments for maximum frequency accuracy.
Response: "0"
the adjustments have been performed successfully.
Query only
See
"Adjust Synthesis"
string
*RST:
force
0 | 1
CAL:LEV:MEAS?
starts adjustments for maximum level accuracy.
Response: "0"
adjustment has been performed successfully.
Query only
See
"Adjust Level"
on page 94
FACTory | CUSTomer
*RST:
FACTory
Remote Control Commands
CALibration Subsystem
on page 205