5.2.2 Default Instrument Settings - Preset Key

The PRESET key performs a defined instrument setup. All parameters and switching
states are preset (also those of inactive operating modes). The default instrument set-
tings provide a reproducible initial basis for further settings.
However, functions concerning the integration of the instrument in a measurement
setup are not changed, for example the GPIB address or reference oscillator settings.
When the instrument is switched on, it is not the preset state that is active, but rather
the instrument state that was set before the instrument was switched on.
User-defined instrument states can be accessed and stored in the "File" menu.
Resetting the instrument to the factory state is possible with the
tion.
Preset
Presets all parameters and switching states.
The following list gives an overview of the presets for the most important generator set-
tings. The other presets can be found in the preset tables of the individual menus and
the information accompanying the remote commands.
"RF frequency" = 1 GHz
"RF level" RF output switched off
"Level" = 30 dBm for instruments including an attenuator
"Level" = -5 dBm for instruments with no attenuator
"Offsets" = 0
"Modulations State" = Off
Uninterrupted level settings are switched off
"Level Attenuator Mode" = AUTO
Internal level control "Level ALC" = AUTO
User correction "Level Ucor" = OFF
"LF output State" = Off
"Sweep State" = Off
Settings that are not affected by the PRESET key
Reference frequency settings ("Ref Oscillator" menu)
Power on settings ("Level/EMF" menu)
Network settings ("Setup" menu)
GPIB address ("Setup" menu)
*IDN? Identification and emulation ("Setup" menu)
Password and settings protected by passwords ("Setup" menu)
Start/Stop Display Update ("Setup" menu)
Display and keyboard settings ("Setup" menu)
Remote command:
*RST

5.2.3 General Configuration of Instrument - Setup Key

Operating Manual 1407.0806.32 ─ 21