Example:
Manual operation:
[:SOURce<hw>]:LIST:INDex:STARt <Start>
Sets the start index of the index range which defines a subgroup of frequency/level
value pairs in the current list. Only the values in the set index range
(:LIST:INDex:STARt ... :LIST:INDex:STOP) are processed in List mode.
Parameters:
<Start>
Example:
Manual operation:
[:SOURce<hw>]:LIST:INDex:STOP <Stop>
Sets the stop index of the index range which defines a subgroup of frequency/level
value pairs in the current list. Only the values in the set index range
(:LIST:INDex:STARt ... :LIST:INDex:STOP) are processed in list mode.
Parameters:
<Stop>
Operating Manual 1407.0806.32 ─ 21
LIST:SEL '/var/list3'
selects list3 for use in List mode.
FREQ:MODE LIST
activates List mode. List3 is processed.
LIST:MODE STEP
selects manual, step-by-step processing of the list.
LIST:IND 5
the frequency/level value pair with index 5 is executed.
TRIG:LIST:SOUR SING
selects triggering by means of the single trigger. The list is exe-
cuted once.
SOUR:LIST:TRIG:EXEC
triggers the processing of the selected list.
See
"Current Index - List Mode"
integer
Range:
0 to list length
*RST:
0
LIST:SEL '/var/list3'
selects list3 for use in List mode.
LIST:IND:STAR 25
sets 25 as start index of the index range.
LIST:IND:STOP 49
sets 49 as stop index of the index range.
FREQ:MODE LIST
activates List mode. The frequency/level value pairs from index
25 to index 49 in list3 are processed. All other entries of the list
are ignored.
See
"List Range In - List Mode"
integer
Range:
0 to list length
*RST:
0
Remote Control Commands
SOURce Subsystem
on page 191
on page 193