R&S
SMB100A
Example:
Usage:
Options:
[:SOURce<hw>]:PULM:TRAin:REPetition <Repetition>
Sets the number of repetitions for each ontime/offtime value pair.
*RST does not affect data lists.
Tip:"0" ignores the corresponding value pair in the pulse train. Thus, you can individu-
ally omit value pairs without deleting them from the table.
Parameters:
<Repetition>
Example:
Options:
Manual operation:
[:SOURce<hw>]:PULM:TRAin:REPetition:POINts?
Queries the length (in points) of the repetition component of the selected list.
Return values:
<Points>
Operating Manual 1407.0806.32 ─ 21
MMEM:CDIR '/var/user/Lists'
selects the directory for the pulse train files.
PULM:TRA:SEL 'P_INCR'
selects P_INCR for editing. P_INCR is created if it does not yet
exist.
PULM:TRA:ONT:POIN?
queries the number of frequency values in P_INCR
Response: 7
P_INCR has 7 ontime entries.
Query only
R&S SMB-K27 (Pulse Train)
Repetition#1{, Repetition#2, ...}
Range:
0...65535
MMEM:CDIR '/var/user/Lists'
selects the directory for the pulse train files.
PULM:TRA:SEL 'P_INCR'
selects P_INCR for editing. P_INCR is created if it does not yet
exist.
PULM:TRA:ONT 10ns,30ns,40ns,...
specifies the ontime values in P_INCR. If the list already con-
tains data, it is overwritten.
PULM:TRA:OFFT 10ns,30ns,40ns,...
specifies the offtime values in P_INCR. If the list already con-
tains data, it is overwritten.
PULM:TRA:REP 1,8,3,...
specifies the number of repetitions for each value pair.
R&S SMB-K27 (Pulse Train)
See
"Edit Pulse Train Data"
integer
Range:
0 to INT_MAX
*RST:
0
Remote Control Commands
SOURce Subsystem
on page 232
397