Adjust Level
Performs all adjustments which affect the level. The acquired correction values
improve the settling time and the signal quality.
Remote command:
:CALibration<hw>:LEVel[:MEASure]?
Adjust Stereo Coder
Performs all adjustments which affect the analog channels of the stereo coder. For the
adjustment of the S/P DIF see Service Manual, chapter 2, "Adjustment".
Remote command:
[:SOURce]:CALibration:STEReo:ANALog[:MEAS]?
5.2.3.2

Hardware Config

In the "Hardware Config" dialog, the installed assemblies together with their variants
and revision states can be displayed for servicing purposes.
To open the "Hardware Config" dialog, select "System" and press the SETUP or
MENU key.
Section "Counter" in the upper part of the menu shows the "Operation Time" (in hours)
and the number of power-on ("Power On Counter").
The second part of the menu is a table that lists the installed assemblies. It is divided
into the sections:
"Common Assembly"
"RF Assembly"
"Baseband Assembly"
Operation Time / h
Displays the operation time in hours.
Remote command:
:DIAGnostic:INFO:OTIMe?
Operating Manual 1407.0806.32 ─ 21
on page 287
on page 289
Instrument Function
General Instrument Settings
on page 288
94