[:SOURce<hw>]:LIST:SELect <Filename>
Selects the specified list. If a new list is to be created, the name can be entered here.
The list is created if it does not yet exist. The list selected here is available for the fur-
ther processing steps (editing) and is used in the instrument when the list mode is acti-
vated.
The files are stored with the fixed file extensions *.lsw in a directory of the user's
choice. The directory applicable to the command is defined with the command
MMEMory:CDIR. A path can also be specified in which case the list mode file in the
specified directory is selected.
*RST does not affect data lists.
Parameters:
<Filename>
Example:
Manual operation:
[:SOURce<hw>]:LIST:TRIGger:EXECute
Starts the processing of a list in list mode. It corresponds to the manual-control com-
mand "Execute Single."
Example:
Usage:
Manual operation:
[:SOURce<hw>]:LIST:TRIGger:SOURce <Source>
Selects the trigger source processing lists.
The names of the parameters correspond to those under sweep mode. SCPI uses
other names for the parameters; these names are also accepted by the instrument.
The SCPI names should be used if compatibility is an important consideration. An
overview of the various names is given in the following table:
R&S name
AUTO
SINGle
EXTernal
Operating Manual 1407.0806.32 ─ 21
'<list name>'
LIST:SEL '/var/list3'
selects list3 for editing.
See
"List Mode Data... - List Mode"
SOUR:LIST:TRIG:EXEC
triggers the processing of the selected list.
Event
See
"Execute Single - List Mode"
SCPI name
IMMediate
BUS
EXTernal
Remote Control Commands
SOURce Subsystem
on page 191
on page 190
Command under manual con-
trol
MODE AUTO
MODE SINGLE or STEP
MODE EXT TRIG SINGLE or
EXT TRIG STEP