:OUTPut<hw>:IMPedance?
Queries the impedance of the RF outputs. It enables you to convert the output level
units between V and W. The impedances cannot be changed.
Return values:
<Impedance>
Example:
Usage:
:OUTPut<hw>:PROTection:CLEar
Resets the protective circuit after it has been tripped. The state of the output is again
determined by OUTPut:STATe.
Example:
Usage:
Manual operation:
:OUTPut<hw>:PROTection:TRIPped?
Queries the state of the protective circuit.
Return values:
<Tripped>
Example:
Usage:
Manual operation:
:OUTPut<hw>[:STATe] <State>
Activates and deactivates the RF output signal (RF ON / RF OFF).
Operating Manual 1407.0806.32 ─ 21
G1K | G50 | G10K
*RST:
G50
Default unit: Ohm
OUTP:IMP
queries the impedance of RF output.
Response: 50
the impedance is 50 ohms
Query only
OUTP:PROT:CLE
resets the protective circuit for RF output.
Event
See
"Overload"
on page 166
0 | 1 | OFF | ON
*RST:
0
OUTP:PROT:TRIP
Queries the state of the protective circuit for RF output A.
Response: 0
The protective circuit has not tripped.
Response: 1
The protective circuit has tripped.
Query only
See
"Overload"
on page 166
Remote Control Commands
OUTPut Subsystem
314