R&S SMB100A Operating Manual page 384

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Parameters:
<Power>
Example:
Manual operation:
[:SOURce<hw>]:LIST:POWer:POINts?
Queries the length (in points) of the LEVel part of the selected list.
Return values:
<Points>
Example:
Usage:
[:SOURce<hw>]:LIST:RESet
Resets the list to the starting point.
Example:
Usage:
Manual operation:
Operating Manual 1407.0806.32 ─ 21
<Power#1>{, <Power#2>, ...} | block data
The data can be given either as a list of numbers (list can be of
any length and list entries must be separated by commas) or as
binary block data. When block data is transferred, 8 bytes are
always interpreted as a floating-point number with double accu-
racy (see
:FORMat[:DATA]
Range:
Minimum level to Maximum level
Default unit: dBm
LIST:SEL '/var/list3'
selects list3 for editing. The R&S SMB generates a new file
automatically, if it does not exist yet.
LIST:POW 0dBm, 2dBm, 2dBm, 3dBm,..
specifies the level values in list3. The number of level values
must correspond to the number of frequency values. The previ-
ous data is overwritten.
See
"Edit List Mode Data... - List Mode"
integer
Range:
0 to INT_MAX
*RST:
0
LIST:SEL '/var/list3'
selects list3 for editing. The R&S SMB generates a new file
automatically, if it does not exist yet.
LIST:POW:POIN?
queries the number of levels in the list file
Response: 327
Query only
LIST:RES
resets the list to the starting point.
Event
See
"Reset - List Mode"
Remote Control Commands
SOURce Subsystem
on page 294).
on page 192
on page 190
369