Sense<Ch>[:Power]:Type; Sense<Ch>[:Power]:Zero; Sense<Ch>:Unit[:Power] - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
SENSe<ch>[:POWer]:TYPE?
The command queries the type of sensor. The type is automatically detected.
Return values:
<Type>
Example:
Usage:
Manual operation:
:SENSe<ch>[:POWer]:ZERO
The command activates the autozero function. Zeroing is required in regular interval (at
least once a day) and if the temperature has varied more than about 5 °C, if the sensor
has been replaced or if measurements of signals with very low power are to be per-
formed. The RF power source must be switched off or disconnected from the sensor
before starting the autozero function.
Example:
Usage:
Manual operation:
:SENSe<ch>:UNIT[:POWer] <Power>
The command selects the unit used for result query with command READ. The power
sensor provides the measured value in Watt. In which unit the measured value is
returned is selected here and might be either Watt, dBm or dBuV.
Parameters:
<Power>
Example:
Manual operation:
Operating Manual 1407.0806.32 ─ 21
string
SENS:TYPE?
queries the type of sensor connected to the POWER SENSOR
connector.
Response: NRP-Z21
the R&S NRP-Z21 sensor is used.
Query only
See
"Current Sensors"
SENS:ZERO
activates autozero function.
Event
See
"Zero - Power Sensors"
DBM | DBUV | WATT
*RST:
DBM
SENS2:UNIT DBM
selects unit dBm for the measured value returned by command
READ.
READ2?
Response: 7.34
7.34 dBm are measured by sensor 2.
See
"Unit"
on page 169
Remote Control Commands
SENSe, READ and INITiate Subsystems
on page 98
on page 156