R&S SMB100A Operating Manual page 365

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Note:
If signal source "Internal" is set, the instrument performs the analog modulations
(AM/FM/ɸM/PM) with this freqeuncy.
In sweep mode
sweep frequency.
Parameters:
<Frequency>
Example:
Manual operation:
[:SOURce<hw>]:LFOutput:FREQuency:MANual <Manual>
Determines the frequency and triggers the next sweep step manually in
LFO:SWE[:FREQ]:MODE
Note: You can select any frequency within the setting range. The range is defined with
LFO:FREQ:STARt
Parameters:
<Manual>
Example:
Manual operation:
[:SOURce<hw>]:LFOutput:FREQuency:MODE <Mode>
Sets the instrument operating mode, and determines the commands to be used for fre-
quency settings.
Operating Manual 1407.0806.32 ─ 21
(LFO:FREQ:MODE
SWEep), the frequency is coupled with the
float
Range:
full frequency range
Increment: see the data sheet: Modulation sources > Resolu-
tion of frequency setting
*RST:
1000
LFO2:FREQ 5kHz
sets the frequency of the LF generator 2 signal to 5 kHz.
See
"LF Gen Freq"
MAN, and
LFO:SWE:[FREQ]:MODE
and LFO:FREQ:STOP.
float
Range:
full frequency range
Increment: see the data sheet: Modulation sources > Internal
modulation generator > Resolution of frequency
setting
*RST:
1000
LFO:SWE:MODE MAN
sets the "Step" sweep mode.
LFO:FREQ:MAN 5 kHz
sets an LF frequency of 5 kHz for the next step in the "Step"
sweep mode.
LFO:FREQ:MODE SWE
sets the LF Sweep mode. An LF frequency of 5 kHz is output.
LFO:FREQ:MAN 5.1 kHz
triggers the next sweep step with a frequency of 5.1 kHz.
See
"Current Freq"
Remote Control Commands
on page 200
STEP.
on page 224
SOURce Subsystem