Note: You can select any frequency within the setting range. The range is defined with
the parameters
FREQuency:STOP. A defined offset and the multiplier factor affect the sweep range
and therefore all correlated parameters. The set frequencies are only absolute values,
if the offset = 0 and the multiplication factor = 1. The multiplier multiplies the frequen-
cies accordingly, and the offset ≠ 0 shifts the frequencies corresponding to the set
value.
f
STARt
Parameters:
<Manual>
Example:
Example:
Manual operation:
[:SOURce<hw>]:FREQuency:MODE <Mode>
Selects the frequency mode for the generating the RF output signal. The selected
mode determines the parameters to be used for further frequency settings.
Operating Manual 1407.0806.32 ─ 21
[:SOURce<hw>]:FREQuency:STARt
* f
+ f
... f
MULTiplier
OFFset
STOP
float
Range:
Increment: see the data sheet: RF characteristics > Resolution
SWE:MODE MAN
sets the Step sweep mode.
FREQ:MODE SWE
sets the frequency sweep mode. The sweep start frequency is
output.
FREQ:MAN UP
triggers the next higher sweep step.
FREQ:MAN 500MHz
outputs 500 MHz RF frequency (must e within the sweep fre-
quency range).
FREQ:MAN DOWN
triggers the next lower sweep step relative to 500 MHz.
See
"Current Freq - Frequency Sweep"
* f
+ f
MULTiplier
OFFset
full frequency range
of setting
Remote Control Commands
SOURce Subsystem
and
[:SOURce<hw>]:
on page 179
343