System:communicate:serial:resource; System:communicate:serial:baud; System:communicate:serial:parity - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Example:
Usage:
Manual operation:

:SYSTem:COMMunicate:SERial:RESource?

Queries the visa resource string for the serial remote control interface. This string is
used for remote control of the instrument.
Return values:
<Resource>
Example:
Usage:
Manual operation:
:SYSTem:COMMunicate:SERial:BAUD <Baud>
Sets the baudrate for the serial remote control interface.
Parameters:
<Baud>
Example:
Manual operation:
:SYSTem:COMMunicate:SERial:PARity <Parity>
Sets the parity for the serial remote control interface.
Parameters:
<Parity>
Example:
Manual operation:
Operating Manual 1407.0806.32 ─ 21
SYSTem:COMMunicate:USB:RESource?
queries the VISA resource string for remote control via the USB
interface.
Response: "USB::72::000000::INSTR"
Query only
See
"Visa Resource Strings"
string
SYSTem:COMMunicate:SERial:RESource?
queries the VISA resource string.
Response: "ASRL1::INSTR"
Query only
See
"Visa Resource Strings"
2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200
*RST:
115200

SYSTem:COMMunicate:SERial:BAUD 115200

determines 115200 baudrate.
See
"Baud Rate"
on page 109
NONE | ODD | EVEN
*RST:
NONE
SYST:COMM:SER:PAR NONE
selects parity NONE.
See
"Parity"
on page 109
Remote Control Commands
SYSTem Subsystem
on page 110