Example:
Usage:
Manual operation:
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER] <User>
The command selects the filter length for user filter mode (SENSe:POWer:FILTer:
TYPE USER). As the filter length works as a multiplier for the time window, a constant
filter length results in a constant measurement time. Values 1 and 2^n are settable.
The time window is fixed to 20 ms.
Parameters:
<User>
Example:
Manual operation:
:SENSe<ch>[:POWer]:FILTer:NSRatio <NSRatio>
The command defines the noise content for fixed noise filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio). This value determines the
proportion of intrinsic noise in the measured result.
Parameters:
<NSRatio>
Example:
Manual operation:
Operating Manual 1407.0806.32 ─ 21
SENS1:FILT:TYPE AUTO
selects auto filter mode for the power sensor connected to the
SENSOR connector.
SENS1:FILT:LENG:AUTO?
queries the automatically set filter length.
Response: 1024
Query only
See
"Filter Length"
on page 172
float
Range:
1 to 65536
*RST:
1
SENS:FILT:TYPE USER
selects user filter mode.
SENS:FILT:LENG 16
sets a filter length of 16. The resulting measurement time is 640
ms (2x16x20 ms).
See
"Filter Length"
on page 172
float
Range:
0.001 to 1
Increment: 0.001
*RST:
0.01
SENS1:FILT:TYPE NSR
selects fixed noise filter mode for the power sensor connected to
the SENSOR connector.
SENS1:FILT:NSR 0.2
sets a noise content of 0.2.
See
"Noise Content"
on page 172
Remote Control Commands
SENSe, READ and INITiate Subsystems
319