Smz Info Update - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Tip: Click on a sensor to get quick information about the firmware version and whether
this sensor measures the peak of the signal.
Remote command:
SENSe<ch>[:POWer]:TYPE?
SENSe<ch>[:POWer]:SVERsion?
SENSe<ch>[:POWer]:SNUMber?
Update
Section "Update" provides access to the file system in order to select a file for an R&S
NRP sensor update (Button "Select NRP File"), the selected file is indicated to the left
of the button. On the left side, the sensor to be updated is selected.
Button "Run Update" starts the update.
Note: If the update is interrupted for example by accidental removal of the sensor dur-
ing the process, the button "Rescue" appears. Thus, you can restart the update proc-
ess.
Prerequisite is that no other sensor is connected to the instrument.
Refer to
rupted update of an R&S NRP sensor"
Remote command:
n.a.
5.2.3.6

SMZ Info Update

The "SMZ Info/Update" dialog covers information on a connected frequency multiplier,
like type, serial number revision state, frequency range and firmware version. You can
directly perform an update of the multiplier firmware.
To access the "SMZ Info/Update" dialog, perform one of the following
On the front panel, press the SETUP key and select "SMZ > SMZ Info Update".
In the block diagram, select "RF Block > Configure > SMZ Info Update"
Operating Manual 1407.0806.32 ─ 21
"How to update an R&S NRP sensor"
on page 324
on page 323
on page 322
on page 98 and
on page 98 for detailed instructions.
Instrument Function
General Instrument Settings
"How to restart an inter-
99