R&S SMB100A Operating Manual page 366

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Parameters:
<Mode>
Example:
Manual operation:
[:SOURce<hw>]:LFOutput:FREQuency:STARt <Start>
Sets the start frequency for the LF sweep.
Parameters:
<Start>
Example:
Manual operation:
[:SOURce<hw>]:LFOutput:FREQuency:STOP <Stop>
Sets the stop frequency for the LF sweep.
Operating Manual 1407.0806.32 ─ 21
CW | FIXed | SWEep
CW|FIXed
Sets the CW frequency mode. CW and FIXed are synonyms.
The instrument operates at a fixed frequency. To set the LF out-
put frequency, use the command
LFOutput<ch>:FREQuency.
SWEep
Sets the sweep mode.
The instrument processes the frequency settings in defined
sweep steps. To determine the corresponding frequency values,
use the commands
[:SOURce<hw>]:LFOutput:FREQuency:
STARt,
[:SOURce<hw>]:LFOutput:FREQuency:STOP
SOURce<hw>]:LFOutput:FREQuency:MANual.
*RST:
CW
LFO:FREQ:MODE SWE
sets the sweep mode.
See
"State"
on page 221
float
Range:
full frequency range
Increment: see the data sheet: Resolution of frequency setting
*RST:
1 KHz
RST*
activates all presettings.
LFO:SWE:MODE AUTO
TRIG0:SOUR SING
LFO:FREQ:STAR 1 kHz
LFO:FREQ:STOP 10 kHz
LFO:FREQ:MODE SWE
LFO:SWE:EXEC
the instrument generates a single sweep cycle from 1 kHz to 10
kHz automatically after a manual trigger event occurs
(:LFOutput:SWEep:EXECute or *TRG). The step width is 1
kHz linear, with 15 ms dell time until the signal switches to the
subsequent step.
See
"Start Freq"
on page 224
Remote Control Commands
SOURce Subsystem
[:SOURce]:
or
[:
351