R&S SMB100A Operating Manual page 344

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
The amplitude can also be linearized automatically by means of a R&S NRP power
sensor connected to the generator output signal. With the aid of the command
SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe, a list
with correction values for external test assemblies can be automatically determined,
e.g. for compensating the frequency response of cables. The correction values can be
acquired any time irrespective of the modulation settings of the generator.
[:SOURce]:CORRection:CSET:CATalog?
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency:POINts?
[:SOURce<hw>]:CORRection:CSET:DATA:POWer
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:POINts?
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
[:SOURce]:CORRection:CSET:DELete
[:SOURce<hw>]:CORRection:DEXChange:AFILe:CATalog?
[:SOURce<hw>]:CORRection:DEXChange:AFILe:EXTension
[:SOURce<hw>]:CORRection:DEXChange:AFILe:SELect
[:SOURce<hw>]:CORRection:DEXChange:AFILe:SEParator:COLumn
[:SOURce<hw>]:CORRection:DEXChange:AFILe:SEParator:DECimal
[:SOURce<hw>]:CORRection:DEXChange:EXECute
[:SOURce<hw>]:CORRection:DEXChange:MODE
[:SOURce<hw>]:CORRection:DEXChange:SELect
[:SOURce<hw>]:CORRection:CSET[:SELect]
[:SOURce<hw>]:CORRection[:STATe]
[:SOURce<hw>]:CORRection:VALue?
[:SOURce<hw>]:CORRection:ZERoing:STATe
[:SOURce]:CORRection:CSET:CATalog?
Requests a list of user correction tables. The individual lists are separated by commas.
The lists are stored with the fixed file extensions *.uco in a directory of the user's
choice. The directory applicable to the commands is defined with the command
MMEMory:CDIR.
Return values:
<Catalog>
Example:
Usage:
Manual operation:
Operating Manual 1407.0806.32 ─ 21
.........................................................................329
............................................................................ 331
................................................................... 335
............................................................................. 336
............................................................................. 336
................................................................. 336
string
MMEM:CDIR '/var/user/ucor'
selects the directory for the user correction files.
CORR:CSET:CAT?
queries which correction tables are available.
Response:UCOR1,UCOR2,UCOR3
the correction tables UCOR1, UCOR2 and UCOR3 are availa-
ble.
Query only
See
"Directory, File List and File Name"
Remote Control Commands
SOURce Subsystem
.................................................... 330
....................................... 330
........................................................... 330
.............................................. 331
.............................................. 332
.............................................332
.................................................. 332
................................333
................................333
.........................................................334
............................................................ 334
............................................................335
on page 127
[:
........................331
329

Advertisement

Table of Contents
loading

Table of Contents