Example:
Manual operation:
[:SOURce<hw>]:LIST:LEARn
Learns the selected list to determine the hardware setting for all list entries. The results
are saved with the list. When the list is activated the first time, these settings are calcu-
lated automatically.
Example:
Usage:
Manual operation:
[:SOURce<hw>]:LIST:MODE <Mode>
Selects how the list is to be processed (similar to SOURce:SWEep:MODE).
Parameters:
<Mode>
Example:
Manual operation:
[:SOURce<hw>]:LIST:POWer <Power>
Fills the Level part of the selected list with data.
*RST does not affect data lists.
Operating Manual 1407.0806.32 ─ 21
see
[:SOURce<hw>]:LIST:INDex:STARt
See
"List Range In - List Mode"
LIST:SEL '/var/list3'
selects list file. The file is created if it does not yet exist.
LIST:LEAR
starts learning of the hardware setting for list3 and stores the
setting.
Event
See
"Learn List Mode Data... - List Mode"
AUTO | STEP
AUTO
Each trigger event triggers a complete list cycle. Possible trigger
settings for :LIST:TRIGger:SOURce are AUTO, SINGle and
EXT.
STEP
Each trigger event triggers only one step in the list processing
cycle. The list is processed in ascending order.
*RST:
AUTO
LIST:MODE STEP
selects step-by-step processing of the list.
See
"Mode - List Mode"
Remote Control Commands