[:Source<Hw>]:Sweep[:Frequency]:Step:logarithmic - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Example:
Manual operation:
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic <Logarithmic>
Sets a logarithmically determined sweep step size for the RF frequency sweep. It is
expressed in percent and you must enter the value and the unit PCT with the com-
mand.
The frequency is increased by a logarithmically calculated fraction of the current fre-
quency according to:
step_log
f
= f
n+1
with f
This parameter correlates with the number of steps
sweep range as follows:
freq_points = ((log f
If you change the step size, the number of steps changes accordingly. The sweep
range remains the same.
Parameters:
<Logarithmic>
Example:
Manual operation:
Operating Manual 1407.0806.32 ─ 21
FREQ:STAR 1GHz
sets the start frequency to 1 GHz.
FREQ:STOP 5GHz
sets the stop frequency to 5 GHz.
SWE:SPAC LIN
sets linear sweep spacing.
SWE:STEP 2 MHz
sets the step width for linear sweep spacing to 2 MHz (RF
sweep) at the RF output. The number of sweep steps for linear
sweep spacing (POINts) is automatically set to 2001.
See
"Step Lin/Log - Frequency Sweep"
= f
+ step_log
n+1
n
n
+ step_log
n
n+1
< f
and n = number of sweep steps
STARt
STOP
- log f
STOP
float
Range:
Increment: 1E-3
*RST:
FREQ:STAR 1GHz
sets the start frequency to 1 GHz.
FREQ:STOP 5GHz
sets the stop frequency to 5 GHz.
SWE:SPAC LOG
sets logarithmic sweep spacing.
SWE:STEP:LOG 10PCT
sets the step width for logarithmic sweep spacing to 10% of the
previous frequency in each instance (for a frequency sweep).
See
"Step Lin/Log - Frequency Sweep"
x f
n
) / log step_log) + 1
STARt
0.01 to 100
1
Remote Control Commands
SOURce Subsystem
on page 180
SWE:FREQ:POIN
within the
on page 180
424