Output<Hw>:Afixed:range:lower; Output<Hw>:Afixed:range:upper; Output<Hw>:Alc:search:mode - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
:OUTPut<hw>:AFIXed:RANGe:LOWer?
Queries the minimum level which can be set when the attenuator is fixed, see
OUTPut<hw>:AMODe.
Return values:
<Lower>
Example:
Example:
Usage:
Manual operation:
:OUTPut<hw>:AFIXed:RANGe:UPPer?
Queries the maximum level which can be set when the attenuator is fixed, see
OUTPut<hw>:AMODe.
Return values:
<Upper>
Example:
Example:
Usage:
Manual operation:
:OUTPut<hw>:ALC:SEARch:MODE <Mode>
Activates/deactivates the RF output during the power search.
Parameters:
<Mode>
Example:
Manual operation:
Operating Manual 1407.0806.32 ─ 21
float
Increment: 0.01
OUTP:AFIX:RANG:LOW?
queries the minimum level for the FIXed setting.
Response: -50
The minimum level is -50 dBm.
Query only
See
"Fixed Range (PEP) In"
float
Increment: 0.01
OUTP:AFIX:RANG:UPP?
queries the maximum level for the FIXed setting for the RF out-
put.
Response: -27
The maximum level is -27 dBm.
Query only
See
"Fixed Range (PEP) In"
NORMal | MINimum
*RST:
NORMal
POW:ALC:SEAR:MODE NORM
during the power search, the RF output is active.
See
"RF During Power Search - ALC"
Remote Control Commands
OUTPut Subsystem
on page 149
on page 149
on page 153
:
: