Return values:
<ErrorCode>
Usage:

*WAI

Wait to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
*OPC).
Usage:

7.3 Preset Commands

The preset commands are not bundled in one subsystem. Therefore, they are listed
separately in this section. In addition, a specific preset command is provided for each
digital standard and for the fader. These specific commands are described in the asso-
ciated subsystems.
Four presetting actions are available:
Activating the default state of all internal instrument functions
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. GPIB address or reference oscillator source settings.
Activating the preset state of the parameters related to the selected signal path
SOURce<hw>:PRESet
Activating the preset state of all parameters that are not related to the signal path
(:DEVice:PRESet
Activating the original state of delivery (factory reset,
on page 285). Only functions that are protected by a password remain unchanged
as well as the passwords themselves.
:DEVice:PRESet
Presets all parameters which are not related to the signal path, including the LF gener-
ator.
Example:
Usage:
Operating Manual 1407.0806.32 ─ 21
integer > 0 (in decimal format)
An error occurred.
(For details see the Service Manual supplied with the instru-
ment).
0
No errors occurred.
Query only
Event
on page 285)
on page 284)
Presets all instruments settings that are not related to the signal
path
Event
Remote Control Commands
Preset Commands
(*RST
on page 282).
:SYSTem:FPReset
(: