Afq Commands - Setup Section; Calibration Subsystem - R&S AFQ100A Operating Manual

I/q modulation generator
Hide thumbs Also See for AFQ100A:
Table of Contents

Advertisement

R&S AFQ100A

AFQ Commands - Setup Section

Calibration Subsystem

The Calibration system contains the commands for adjustment. Adjustment is triggered by the query
commands. The response "0" indicates error-free adjustment, and the response "1" means that an error
occurred during adjustment.
Command
CALibration:ALL[:MEASure]?
CALibration:AOUTput<CH>:AMPLitude?
CALibration:AOUTput<CH>:DELay?
CALibration:ALL[:MEASure]?
The command starts all internal adjustments for which no external measuring equipment is needed.
"CAL:ALL:MEAS?"
Example:
Response: "0"
*RST value
-
CALibration:AOUTput<CH>:AMPLitude?
The command adjusts the amplitude of the (selected) analog output.
The instrument generates a series of signal levels and converts them using a precise ADC. The
procedure takes about 30 seconds.
The command is a query command and therefore has no *RST value.
Example:
CAL:AOUT:AMPL?
Response: '0'
*RST value
-
1401.3084.32
'starts the adjustment of all functions for the entire instrument.
'adjustment has been performed successfully.
Resolution
Options
-
'starts all internal adjustments for which no external measuring equipment is
needed. The adjustments with external measuring equipment are described in
the service manual.
'calibration was successful
Resolution
Options
-
AFQ Commands - Setup Section
Parameters
6.72
Default
Remark
unit
Query only
Query only
SCPI
Compliant
SCPI
Device-specific