:SYSTem:COMMunicate:SERial:SBITs <SBits>
Sets the number of stop bits for the serial remote control interface.
Parameters:
<SBits>
Example:
Manual operation:

:SYSTem:COMMunicate:SOCKet:RESource?

Queries the visa resource string for remote control via LAN interface, using TCP/IP
socket protocol.
Return values:
<Resource>
Example:
Usage:
Manual operation:
:SYSTem:IDENtification <Identification>
Selects the mode the instrument identification is performed.
Parameters:
<Identification>
Example:
Manual operation:
:SYSTem:LANGuage <Language>
Sets the remote control command set.
The instrument can also be remote controlled via the command set of several other
generators, for example HP generator. See the Application Note
load area of the product site on the Internet.
Operating Manual 1407.0806.32 ─ 21
1 | 2
*RST:
1
SYST:COMM:SER:SBIT 2
selects 2 stop bits.
See
"Stop Bits"
on page 110
string
SYSTem:COMMunicate:SOCKet:RESource?
Response: "TCPIP::10.113.1.150::5025::SOCKET"
Query only
See
"Visa Resource Strings"
AUTO | USER
AUTO
The "IDN String" and the "OPT String" are set automatically.
USER
Enables the selection of user definable "IDN String" and "OPT
String".
*RST:
AUTO
SYST:IDEN USER
selects the user defined identification string.