[:Source<Hw>]:Sweep[:Frequency]:Spacing; [:Source<Hw>]:Sweep[:Frequency]:Step[:Linear] - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Manual operation:
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing <Spacing>
Selects the mode for the calculation of the frequency sweep intervals. The frequency
increases or decreases by this value at each step.
The keyword [:FREQuency] can be omitted. Then the command is SCPI-compliant.
Parameters:
<Spacing>
Example:
Manual operation:
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear] <Linear>
Sets the step size for linear RF frequency sweep steps.
This parameter is related to the number of steps
FREQuency]:POINts) within the sweep range as follows:
f
STARt
freq_points = (f
If you change the step size, the number of steps changes accordingly. The sweep
range remains the same.
The keywords [:FREQuency] and [:LINear] can be omitted. The command is then
SCPI-compliant.
Parameters:
<Linear>
Operating Manual 1407.0806.32 ─ 21
See
"Shape - RF Frequency Sweep"
LINear | LOGarithmic
LINear
With the linear sweep, the step width is a fixed frequency value
which is added to the current frequency. The step width for lin-
ear sweep is entered in Hz (see
FREQuency]:STEP[:LINear]
LOGarithmic
With the logarithmic sweep, the step width is a constant fraction
of the current frequency. This fraction is added to the current fre-
quency. The logarithmic step width is entered in % (see
SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
on page 424).
*RST:
SWE:SPAC LIN
selects linear sweep spacing for a frequency sweep at the RF
output.
See
"Spacing - Frequency Sweep"
< f
STOP
/ step_lin) + 1
SPAN
float
Range:
Increment: see the data sheet: RF characteristics > Resolution
Remote Control Commands
[:SOURce<hw>]:SWEep[:
on page 423).
LINear
([:SOURce<hw>]:SWEep[:
full frequency range
of setting
SOURce Subsystem
on page 179
[:
on page 179
423