Sense<Ch>[:Power]:Filter:sonce; Sense<Ch>[:Power]:Filter:type - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe <MTime>
The command defines the timeout for fixed noise filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio). This value ensures limited
settling times.
Parameters:
<MTime>
Example:
Manual operation:
:SENSe<ch>[:POWer]:FILTer:SONCe
The command activates the search for the optimum filter length for the current mea-
surement conditions. The found filter length can be retrieved with com-
mand :SENSe:POWer:FILTer:LENGth:USER?. This command is only available for
user filter mode (:SENSe:POWer:FILTer:TYPE USER).
Example:
Usage:
Manual operation:
:SENSe<ch>[:POWer]:FILTer:TYPE <Type>
The command selects the filter mode. The filter length is the multiplier for the time win-
dow and thus directly influences the measurement time.
Operating Manual 1407.0806.32 ─ 21
float
Range:
1 to 999.99
Increment: 0.01
*RST:
4
Default unit: s
SENS1:FILT:TYPE NSR
selects fixed noise filter mode for the power sensor connected to
the SENSOR connector.
SENS1:FILT:NSR .2
sets a noise content of 0.2.
SENS1:FILT:NSR:MTIM 5
limits the settling time to 5 seconds
See
"Timeout"
on page 172
SENS:FILT:TYPE USER
selects user filter mode.
SENS:FILT:SONC
activates the search for the optimum filter length.
SENS:FILT:LENG?
returns the found optimum filter length.
Response: 128
Event
See
"Auto Once"
on page 173
Remote Control Commands
SENSe, READ and INITiate Subsystems
320

Advertisement

Table of Contents
loading