Key Definitions

Softkeys displayed by pressing the [Cal] key

Softkeys displayed by pressing the
Table B-34
Calibration (Calibration) menu
Key operation:
[Cal]
Softkey
ADAPTER [ ] ä
Displays Table B-35, "Adapter configuration (Adapter) menu," on page 378. In
[ ] of the softkey label, the currently selected adapter configuration is displayed.
FIXTURE COMPEN ä
Displays Table B-37, "Fixture compensation (Fixture Comp) menu," on page
380.
CAL POINT [ ]
For fixture compensation (FIXTURE COMPEN) and user calibration (USER
CAL), selects a preset single fixed frequency point [FIXED], which does not
change regardless of conditions, or a frequency point determined by the
user-specified sweep range [USER]. This configuration is commonly applied to
both user calibration and fixture compensation.
USER CAL ä
Displays Table B-39, "User calibration (User Cal) menu," on page 381. This key
is valid only when 7-mm 42942A or PROBE 42941A has been selected with the
ADAPTER [ ] key.
PORT EXTENSION ä
Displays Table B-42, "Port extension (Prt Extension) menu," on page 382. This
key is valid only when 7-mm 42942A or PROBE 42941A has been selected with
the ADAPTER [ ] key.
Table B-35
Adapter configuration (Adapter) menu
Key operation:
[Cal] - ADAPTER [ ]
Softkey
NONE
You select this key when using no extension cable from the measurement
|
terminal on the front panel to connect a DUT during four-terminal pair
measurement. This is the proper configuration when using the HP 16047E test
fixture (direct-coupled fixture for devices with leads) and similar devices.
4TP 1M
You select this key when using a 1-m extension cable from the measurement
|
terminal on the front panel to connect a DUT during four-terminal pair
measurement. This is the proper configuration when using the HP 16048G test
cable (4TP 1m cable).
4TP 2M
You select this key when using a 2-m extension cable from the measurement
|
terminal on the front panel to connect a DUT during four-terminal pair
measurement. This is the proper configuration when using the HP 16048H test
cable (4TP 2m cable).
7mm 42942A
You select this key when connecting the HP 42942A Terminal Adapter to the
|
measurement terminal on the front panel.
PROBE 42941A
You select this key when connecting the HP 42941A Impedance Probe to the
measurement terminal on the front panel.
378
[Cal]
Function
Function
key
GPIB command
(N/A)
(N/A)
CALP {FIXED|USER}
(N/A)
(N/A)
GPIB command
E4TP OFF
E4TP M1
E4TP M2
E4TP APC7
E4TP PROBE
Appendix B