Specifications and Supplemental Performance Characteristics

Basic Characteristics

Specifications describe the instrument's warranted performance over the temperature range
of 0°C to 40°C (except as noted). Supplemental performance characteristics are intended to
provide information that is useful for applying non-warranted performance parameters to
the instrument. Such parameters are denoted as SPC (supplemental performance
characteristics), typical, or nominal. Warm-up time must at least 30 minutes after power-on
for all specifications.
Basic Characteristics

Measurement Parameter

Impedance Parameters

Measurement Terminal

Configuration
Connector type

Source Characteristics

Frequency
Range
Resolution
Accuracy
without Option 1D5
with Option 1D5
Voltage Signal Level
Range
Resolution
Accuracy
324
|Z|-q, R-X, Ls-Rs, Ls-Q, Cs-Rs, Cs-Q, Cs-D, |Y|-q,
G-B, Lp-G, Lp-Q, Cp-G, Cp-Q, Cp-D, Complex Z-Y,
|Z|-Ls, |Z|-Cs, |Z|-Lp, |Z|-Cp, |Z|-Rs, |Z|-Q, |Z|-D,
Lp-Rp, Cp-Rp
Four-terminal pair configuration
Four BNC (female) connectors. Can be converted to
one-port terminal using the 42942A Terminal Adapter
(7-mm port) or the 42941A Impedance Probe
(3.5-mm port).
40 Hz to 110 MHz
1 mHz
± 20 ppm (at 23 ± 5°C)
± 40 ppm (at 0 to 55°C)
± 0.13 ppm (at 0 to 55°C)
5 mVrms to 1 Vrms
1 mV