Figure 8-27
Limit line test function ON (example of rejection by test)
NOTE
When the limit line test is turned on, the test result (pass/fail) is output as a PASS/FAIL
signal through the 24-bit I/O port of the rear panel.
Step 4. Operate the
the event that the limit line test indicates failure.
Softkey label
BEEP FAIL ON off
BEEP FAIL on OFF
NOTE
The limit line table can be edited with the limit line test function turned on.
If the limit line test function is turned on, the pass/fail result is indicated by the following
four methods.
"PASS" or "FAIL" is displayed in the Limit Line Test Result field on the right side of
the screen.
If the fail beep setting is On (
failure.
If you execute the list display (
(measured value) by following the procedures in "Print the measurements and settings
(text)" on page 296, an asterisk "*" indicating a failure is displayed next to
measurement points that exceeds the limit.
Chapter 8
Set a limit to the trace and make pass/fail evaluation
(or
BEEP FAIL on OFF
BEEP FAIL ON off
BEEP FAIL ON off
LIST VALUE
Analysis and Processing of Result
) key to activate a beep sound in
Beep sound when rejected
ON (beeps)
OFF (doesn't beep)
), a beep sounds in the event of a
key) of the measurement parameter value
263