Results Of Analysis; Determine The Self-Resonance Frequency And Resonant Impedance - Agilent Technologies 4294A Operating Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

Learning Operation Basics

Results of Analysis

Results of Analysis

Determine the Self-resonance Frequency and Resonant Impedance

Step 1. Press the
Step 2. Press the
identified by number 0) appears in the center of the graph.
Step 3. Press the
should change to
performs a search action upon completion of each sweep cycle.
Step 4. Press the
tracking function is enabled, the instrument detects and moves the main marker to the
minimum |Z| value each time it has completed one sweep cycle.
You can determine the self-resonance frequency and resonant impedance by reading the
sweep parameter (frequency) and measurement parameter values indicated by the main
marker that appears in the upper-left area of the graph for Trace A. In the example shown
in Figure 2-9, the DUT exhibits its self-resonance frequency and resonant impedance at
approximately 5.13 MHz and 155 mW, respectively.
Figure 2-9
Result of Searching for the Self-Resonance Frequency (with the Search Tracking
Function On)
42
key to activate Trace A.
[A]
key to display the Search menu. The main marker (a trace marker
[Search]
key to turn on the search tracking function (the key label
SEARCH TRK on OFF
SEARCH TRK ON off
key to move the main marker to the minimum |Z| value. Since the search
MIN
). When this function is enabled, the instrument
Chapter 2