Learning Operation Basics
Fixture Compensation
Figure 2-3
Setting Up the Test Circuit for SHORT Compensation
Step 2. Press the
measuring the compensation data, a message "WAIT--MEASURING STANDARD" is
displayed in the Parameter Setting field in the upper-left area of the screen. Upon
completion of measurement, the
indicating that the SHORT compensation function is turned on.
Step 3. Remove the short bar from the Agilent 16047E's test electrodes.
NOTE
The Agilent 4294A provides three fixture compensation functions: OPEN, SHORT, and
LOAD. You can turn on/off each of these compensation functions individually and use
them in any combination to perform the fixture compensation process. In typical
measurement use, it is recommended that you execute the OPEN and SHORT
compensation functions.