Agilent Technologies 4294A Operating Manual page 453

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

The received block, equation, or string type program data comply with the standard, but the
amount of data exceeds the limit that the 4294A can deal with, due to memory or
device-specific conditions related to memory.
-210
Trigger error
A trigger error. An error other than -211 to -219 has occurred.
-211
Trigger ignored
A trigger command or trigger signal is received and recognized by the 4294A, but it was
ignored due to the timing relationship with the 4294A (for example, when the 4294A was
not ready to respond).
-113
Undefined header
A header not defied for the Agilent 4294A is received. For example, the header "*XYZ",
which is not defined for the 4294A is received.
19
UNEXPECTED DATA DETECTED
During the measurement of calibration data (OPEN, SHORT, LOAD) for user calibration,
fixture compensation, and four-terminal pair extension setting, abnormal data are detected,
and the acquisition of the calibration data is aborted. Check that there is no problem in
standards or setup you use.
142
WRONG I/O PORT DIRECTION
For a 24-bit input/output (I/O) shared port (port C and port D), you attempt to execute
input/output (I/O) of data in the direction opposite from the preset input/output direction.
The command is thus invalid. For example, if you use the CIN command to specify port C
as an input port and then use the OUTCIO <value> command to output data from port C,
this error is displayed.
Appendix C
U
W
Error messages
Trigger error
445

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents