D. Calibration When an Exclusive Fixture is Connected to
the 42942A
When an exclusive test fixture supplied by Agilent Technologies is connected to the 7-mm
port of the 42942A Terminal Adapter, which is connected to the four-terminal pair port of
the Agilent 4294A, and a device is connected to the test fixture, perform the calibration
procedure described below.
Figure 6-4
Calibration When an Exclusive Fixture is Connected to the 42942A
This calibration procedure is applied when the following Agilent test fixtures are
connected to the 42942A's 7-mm port:
•
16092A Lead/Chip component test fixture
•
16093A Lead component test fixture
•
16093B Lead component test fixture
•
16191A Chip component test fixture
•
16192A Chip component test fixture
•
16193A Chip component test fixture
•
16454A Magnetic material test fixture
Step 1. Make sure that the adapter selection is set to "42942A Terminal Adapter (
Chapter 6
D. Calibration When an Exclusive Fixture is Connected to the 42942A
Calibration
)."
7mm 42942A