error occurs, for example, when data in a file to be read out is destroyed or when an
extension for a file name does not match with the contents of the file.
134
REDUCE OSC LEVEL
The oscillator level is considerably higher than the measured |Z| (absolute value of
impedance) of the DUT, causing increased nonlinear measurement error that cannot be
ignored. This message is displayed when the oscillator level is set to greater than 0.5 V
while |Z| of the DUT is smaller than 48 W. To avoid this error, set the oscillator level below
0.5 V.
110
SAVE ERROR
When saving a file, anomalies in the storage media are detected. For example, if you
attempt to save a file on a floppy disk, it may be damaged. Check that there is no problem
with the storage media.
90
SCALE PARAMETER NOT VALID
You attempt to set a display scale not available for the selected display format and, as a
result, the command is ignored. For example, when the log scale (command: FMT LOGY)
is selected as the display format, you attempt to perform the SCALE/DIV setting
(command: SCAL <value>).
102
SEARCH RANGE TYPE IS NOT SEGMENT
Although in the list sweep all segments (entire sweep range) are specified as the search
range (command: SEARNG FULL), you attempt to execute a command to specify a
certain segment as a search target (SEGMNUM <value>) and, as a result, the command is
ignored. Before executing the search function by specifying a certain list sweep segment,
set the search range to a single segment (command: SEARNG SEGMENT).
101
SEARCH WIDTH OFF
When the band width function is off, you attempt to execute a command to search the
cutoff point on the trace (WIDSIN or WIDSOUT) and, as a result, the command is ignored.
Before executing the cutoff point search of the band width function, turn on the band width
function (command: WIDT ON).
-330
Self-test failed
The result of the self-test is failure. Contact your local Agilent Technologies sales office or
service center, or refer to the service manual.
-221
Setting conflict
A program data element complying with the syntax standard is analyzed but the 4294A
cannot execute it at present.
-150
String data error
An error not described by error numbers from -151 to -159 occurs during the syntax
analysis of a string data element.
-158
String data not allowed
A string data element is received where the 4294A does not accept any string data element.
-130
Suffix error
Appendix C
REDUCE OSC LEVEL
S
Error messages