Preparation of Measurement Accessories
Adapter Setting
Connecting 100 W Resister (Load Data Measurement)
Figure 4-3
Step 8. Press the
NOTE
When the load data measurement is completed, the softkey label changes to
LOAD [DONE]
Step 9. Press the
compensation and load data. The adapter setup data is automatically saved to the EEPROM
(nonvolatile memory) in the Agilent 4294A.
NOTE
Press the
canceled, the previously obtained data is saved as the available adapter setup data.

Adapter Setup Procedure for the 16334A

Perform the adapter setup for the 16334A Tweezers Type Test Fixture in accordance with
the following procedure:
NOTE
Adapter setup should be performed at an ambient temperature as close as possible to the
temperature in which the device measurement will actually be made.
If the ambient temperature for your measurement is outside the range of ±5°C from the
adapter setup temperature, it is recommended that you perform adapter setup again at the
appropriate temperature.
It takes a few minutes to complete each setup data measurement.
During each data measurement, the message "WAIT--MEASURING ADJUSTMENT
KIT" is displayed in the instrument state area in the LCD's upper-left side.
Step 1. Make sure that the 16334A is correctly connected to the four-terminal pair port of the
72
key to start the load data measurement.
LOAD [-]
.
key to start calculating the adapter setup data from the measured phase
done
key to cancel the adapter setup process. If the adapter setup process is
cancel
Chapter 4