Table B-85
Segment edit (Edit Limit) menu (2/3)
Key operation:
[System] - LIMIT TEST - EDIT LIMIT LINE - EDIT - more 1/3
[System] - LIMIT TEST - EDIT LIMIT LINE - ADD - more 1/3
Softkey
DELTA LIMIT
Sets the limit center value of the measured value at the sweep stop point of the
segment
between the upper and lower limits) with the DELTA LIMIT key. This kind of
limit configuration set by the center value and the difference value can be used as
an alternative to the limit configuration set by the upper and lower limits.
more 2/3
Displays Table B-86, "Segment edit (Edit Limit) menu (3/3)," on page 411.
done
Has the same function as the done key in Table B-84, "Segment edit (Edit Limit)
menu (1/3)," on page 409.
cancel
Has the same function as the cancel key in Table B-84, "Segment edit (Edit
Limit) menu (1/3)," on page 409.
1. Even if you have set the limits with the MIDDLE VALUE (center value) key and the DELTA LIMIT (difference value) key, you
can convert the table into the upper and lower limit format by pressing the UPPER LIMIT key or the LOWER LIMIT key. On
the other hand, even if you have set the limits with the UPPER LIMIT (upper limit) key and the LOWER LIMIT (lower limit)
key, you can convert the table into the center and difference value format by pressing the MIDDLE VALUE key or the DELTA
LIMIT key.
Table B-86
Segment edit (Edit Limit) menu (3/3)
Key operation:
[System] - LIMIT TEST - EDIT LIMIT LINE - EDIT - more 1/3 - more 2/3
[System] - LIMIT TEST - EDIT LIMIT LINE - ADD - more 1/3 - more 2/3
Softkey
TEST ON off
Toggles on and off the limit line test for the segment. When the limit line test is
off, * is displayed to the right of the segment number in the completed limit line
table. Note that * is not displayed when the pointer (>) is displayed to the right of
the segment number.
MKR ® START
Substitutes the sweep parameter value at the main marker on the current active
trace (A or B) into the sweep start value of the segment.
MKR ® STOOP
Substitutes the sweep parameter value at the main marker on the current active
trace (A or B) into the sweep stop value of the segment.
MKR ® VALUE STAR
Substitutes the sweep parameter value at the main marker on the current active
trace (A or B) into the limit center point at the sweep start point of the segment.
MKR ® VALUE STOP
Substitutes the sweep parameter value at the main marker on the current active
trace (A or B) into the limit center point at the sweep stop point of the segment.
more 3/3
Displays Table B-84, "Segment edit (Edit Limit) menu (1/3)," on page 409.
done
Has the same function as the done key in Table B-84, "Segment edit (Edit Limit)
menu (1/3)," on page 409.
cancel
Has the same function as the cancel key in Table B-84, "Segment edit (Edit
Limit) menu (1/3)," on page 409.
Appendix B
Softkeys displayed by pressing the [System] key
Function
1
. When setting the center value, also set the difference value (difference
Function
Key Definitions
GPIB command
LIMDSTOP <value>
(N/A)
LIMSDON
LIMSQUI
GPIB command
LIMSTEST {ON|OFF}
MKRLIMSTAR
MKRLIMSTOP
MKRLIMVSTAR
MKRLIMVSTOP
(N/A)
LIMSDON
LIMSQUI
411