Adapter Setup Procedure For The 16451B - Agilent Technologies 4294A Operating Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

Agilent 4294A.
NOTE
The 16334A must be directly connected to the four-terminal pair port of the Agilent
4294A. If another accessory such as a cable or adapter is inserted between the 16334A and
the four-terminal pair port of the Agilent 4294A, the correct adapter setup data cannot be
obtained and correct device measurement cannot be performed after the adapter setup.
Step 2. More than 30 minutes warm-up time is required after turning on the Agilent 4294A.
Step 3. Make the correct adapter selection for the 16334A in accordance with the "Adapter
Selection" on page 68.
Step 4. After performing Step 4 of the "Adapter Selection" on page 68, press the
display the Adapter Setup Menu.
Step 5. Put the device contacts of the 16334A into the OPEN condition. Refer to the operation
manual of the 16334A for instructions on switching to the OPEN condition.
Step 6. Press the
NOTE
The traces on the LCD display do not change during the phase compensation data
measurement.
When the phase compensation data measurement is completed, the softkey label changes
to
PHASE COMP [DONE]
Step 7. Connect a 100 W LOAD device to the 16334A.
Step 8. Press the
NOTE
When the load data measurement is completed, the softkey label changes to
LOAD [DONE]
Step 9. Press the
compensation and load data. The adapter setup data is automatically saved to the EEPROM
(nonvolatile memory) in the Agilent 4294A.
NOTE
Press the
canceled, the previously obtained data is saved as the available adapter setup data.

Adapter Setup Procedure for the 16451B

Perform the adapter setup for the 16451B Dielectric Test Fixture in accordance with the
following procedure:
NOTE
Adapter setup should be performed at an ambient temperature as close as possible to the
temperature in which the device measurement will actually be made.
If the ambient temperature for your measurement is outside the range of ±5°C from the
adapter setup temperature, it is recommended that you perform adapter setup again at the
appropriate temperature.
It takes a few minutes to complete each setup data measurement.
Chapter 4
key to start the phase compensation data measurement.
PHASE COMP [-]
.
key to start the load data measurement.
LOAD [-]
.
key to start calculating the adapter setup data from the measured phase
done
key to cancel the adapter setup process. If the adapter setup process is
cancel
Preparation of Measurement Accessories
Adapter Setting
key to
SETUP
73