A. Calibration When Using Direct Connection Type Test
Fixture
This section describes the calibration procedures to follow when a direct connection type
test fixture is connected to the four-terminal pair port of the Agilent 4294A and a device is
connected to the test fixture contacts.
Figure 6-1
Calibration When Direct Connection Type Test Fixture is Used
Direct connection type test fixtures include:
•
16034E Chip component test fixture
•
16034G Chip component test fixture
•
16034H Chip component test fixture
•
16044A Chip component test fixture
•
16047A Lead component test fixture
•
16047C Lead component test fixture
•
16047D Lead component test fixture
•
16047E Lead component test fixture
•
16314A Balanced/Unbalanced 4-Terminal Converter
Although the following test fixtures have cables, their calibration must be performed in the
same way as a direct connection type test fixture:
•
16065A External DC Bias Fixture
•
16089A Kelvin Clip Lead
Chapter 6
A. Calibration When Using Direct Connection Type Test Fixture
Calibration