Calibration When Using Direct Connection Type Test Fixture - Agilent Technologies 4294A Operating Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

A. Calibration When Using Direct Connection Type Test
Fixture
This section describes the calibration procedures to follow when a direct connection type
test fixture is connected to the four-terminal pair port of the Agilent 4294A and a device is
connected to the test fixture contacts.
Figure 6-1
Calibration When Direct Connection Type Test Fixture is Used
Direct connection type test fixtures include:
16034E Chip component test fixture
16034G Chip component test fixture
16034H Chip component test fixture
16044A Chip component test fixture
16047A Lead component test fixture
16047C Lead component test fixture
16047D Lead component test fixture
16047E Lead component test fixture
16314A Balanced/Unbalanced 4-Terminal Converter
Although the following test fixtures have cables, their calibration must be performed in the
same way as a direct connection type test fixture:
16065A External DC Bias Fixture
16089A Kelvin Clip Lead
Chapter 6
A. Calibration When Using Direct Connection Type Test Fixture
Calibration
133