You can use the following procedure to select how the sweep parameter value is displayed
in the Marker Sweep Parameter Value field: time relative to the sweep start or relaxation
time. This function allows you to read the sweep parameter value in a unit of time at each
point on the trace in the zero span sweep.
Step 1. Press the
Step 2. Press the
Step 3. Select the display method of the marker sweep parameter value.
Display in the Marker Sweep Parameter Value field
(marker X-axis display)
Value of the parameter selected as the sweep parameter
(frequency, oscillator level, or dc bias level)
Time between the sweep start and the completion of the
measurement at the marker position
Relaxation time =
frequency)
NOTE
You can select relaxation time as the method to display the marker sweep parameter value
(marker X-axis) only when the frequency is selected as the sweep parameter.
When you use the D marker, the relative value from the D marker position is displayed in
the Marker Sweep Parameter Value area, regardless of the selection of the marker sweep
parameter value (marker X-axis) display.
Note that in the
X-axis display is displayed as follows.
Softkey label
MKR X-AXIS [STIM]
MKR X-AXIS [TIME]
MKR X-AXIS [1/(2pF)]
Chapter 8
key to display the Utility menu.
[Utility]
key to display the Marker X-Axis menu.
MKR X-ASIX [ ]
1
--------- -
(F: measurement
2pF
of the key label of the
[ ]
Analysis and Processing of Result
Key operation
STIMULUS
TIME
1/(2pF)
key, the currently selected marker
MKR X-AXIS [ ]
Selected marker X-axis display
Sweep parameter
Time
1
--------- -
Relaxation time =
2pF
frequency)
or relaxation time
(F: Measurement
245