Trace-Based Comparison And Calculation; Identifying Differences Between Data And Memory Traces Through Comparison Or Calculation - Agilent Technologies 4294A Operating Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

Trace-based Comparison and Calculation

Traces A and B each provide two different trace types: data and memory. The data trace
keeps track of the DUT's characteristics as it is being tested, while the memory trace
captures the measurement data at a specific point in time and displays that data. With the
data trace, you can perform a simple calculation on the measurement data and display the
results. For example, you can display the differential between the measured data and the
memory trace data or the result of subtracting your specified offset value from the
measured data. Table 7-1 lists the trace types and their functions:
Trace Types
Data trace (yellow
Trace A
Memory trace
(green
Data trace (blue
Trace B
Memory trace
(brown
1. Factory default settings.For how to change this setting, refer to
of Each Screen Item" on page 208
These features add flexibility to your testing task; for example, on the same screen you can
compare the data and memory traces with each other or simultaneously display the results
of a calculation as shown by the data and memory traces.
Identifying Differences between Data and Memory Traces through
Comparison or Calculation
Step 1. Activate the trace for which you want to perform calculation or comparison between the
two trace types.
To activate:
Trace A
Trace B
Step 2. Press the
key to display the Display menu.
[Display]
Step 3. Press the
DEF TRACE [ ]
Step 4. Press the
key to have the data trace directly store and display the measurement data.
DATA
Note that
is selected by default in the preset state.
DATA
Step 5. Start testing the DUT so that the data trace displays the results (for the steps necessary to
perform such a measurement, refer to the appropriate chapters in this guide). You will store
this measurement data into the memory trace later in this procedure.
Chapter 7
Setting Up the Display of Measurement Results

Trace-based Comparison and Calculation

Displays measurement data/calculations
1
)
(updated during measurement cycle)
Stores and displays previously measured data (captures
data when
1
DATA ® MEMORY
)
Displays measurement data/calculations
1
)
(updated during measurement cycle)
Stores and displays previously measured data (captures
1
data when
DATA ® MEMORY
)
.
Press:
[A]
[B]
key to display the Trace Definition menu.
Functions
is pressed)
is pressed)
"Customizing the Color
179

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents