Setting Measurement Conditions
Sweeping Multiple Sweep Ranges with Different Conditions in a Single
Action (List Sweep)
The setting above will add a new value for the dc current bias level to the cell located in the
segment row and the dc bias column (BIAS) of the list sweep table in editing mode.
Step 21. Press the
Step 22. Press the
Step 23. Select the measurement bandwidth for the segment.
Measurement bandwidth
1 (measurement speed maximized)
2
3
4
5 (measurement accuracy maximized)
The setting above will add a new value for the measurement bandwidth to the cell located
in the segment row and the measurement bandwidth column (BW) of the list sweep table in
editing mode.
Step 24. Press the
Step 25. Press the
point averaging in the parameter settings located in the upper left area of the screen.
Step 26. Use the keys or rotary knob of the ENTRY block in one of the following ways to specify a
value for the point averaging.
The setting above will add a new value for the point averaging in the cell located in the
segment row and the point averaging column (AVG) of the list sweep table in editing
mode.
Step 27. When editing of the segment is finished, press the
Pressing this key also returns you to the list sweep menu.
NOTE
If you want to abort editing of the segment and restore the original state before entering the
edit mode, press the
Step 28. With the segment number for the edited segment selected (">" is displayed to the left of the
segment number of the SEG line in the list sweep table), press the PEN key. The number of
the pen selected to draw the trace of the segment is displayed in the Setting Parameter
122
unit key (
,
,
[G/n]
[M/m]
[k/m]
Turn the rotary knob (
m m m m
) until the desired value is set.
Press the step keys ([­][¯]) to set the desired value.
key to display the segment edit menu (3/3).
more 2/3
key to display the measurement bandwidth setting menu.
BANDWIDTH [ ]
key to return to the segment edit menu (3/3).
return
POINT AVG FACTOR
Enter the desired value with the numeric keys (
key.
[´1]
Turn the rotary knob (
) until the desired value is set.
m m m m
Press the step keys ([­][¯]) to set the desired value.
key.
cancel
, or
).
[´1]
Key stroke
1 FAST
2
3
4
5 PRECISE
key. This will display the current setting for the segment's
to
[0]
[9]
key to save the edited segment.
done
, and
) and then press the
[.]
Chapter 5