Key Definitions

Softkeys displayed by pressing the [Trigger] key

Softkeys displayed by pressing the
Table B-56
Trigger (Trigger) menu
Key operation:
[Trigger]
Softkey
HOLD
Selects the sweep stop. The status indication area at the left side of the screen
|
shows "Hld." After the stop of the sweep, if "*" is displayed at the top of the
status indication area, it indicates that the sweep was stopped before completion
and therefore the data trace has not been updated completely. To update the data
trace completely, press the SINGLE key to execute a new single sweep.
SINGLE
Performs a single sweep to update the data trace and then goes into the sweep
|
stop status.
NUMBER OF GROUPS
Sets the sweep group count (the number of sweeps specified by users). When the
|
user finishes entry of the sweep group count with the numeric keys, the sweeps
are started. After the sweeps of the specified count, the sweep stop status occurs.
If the sweep-to-sweep averaging function is on, specify a sweep group count
equal to or larger than the averaging factor (count) to execute averaging
completely. When the group count is entered, the averaging counter is reset to 0
and the sweeps are started. Specify an integer between 0 (no sweep) and 999 for
the sweep group count.
CONTINUOUS
Selects the automatic continuous sweep mode. The trace is updated for each
sweep.
SOURCE [ ] ä
Displays Table B-57, "Trigger source (Trig source) menu," on page 390. In [ ] of
the softkey label, the currently selected trigger source is displayed (internal
trigger: [FREE RUN], external trigger input terminal: [EXTERNAL],
GPIB/LAN: [HP-IB], or manual: [MANUAL]).
Table B-57
Trigger source (Trig source) menu
Key operation:
[Trigger] - SOURCE [ ]
Softkey
FREE RUN
Selects the internal trigger mode (free run), in which triggers are automatically
|
generated inside the 4294A.
EXTERNAL
Selects triggers from the external trigger input terminal (BNC(f) connector) on
|
the rear panel.
HP-IB/LAN
Selects triggers via GPIB or LAN. If this mode is selected, you can generate
|
triggers from the external controller using the *TRG command.
MANUAL
Selects manual triggering. After the selection, pressing the MANUAL key
generates a single trigger.
EVENT [ ]
Selects applying a trigger to each measurement point [ON POINT] or to each
sweep [ON SWEEP]. This configuration is valid for each trigger mode of
EXTERNAL, GPIB/LAN, and MANUAL described above. For FREE RUN,
this key has no meaning.
EXT POLRTY [ ]
Selects the polarity of the trigger signal supplied from the external trigger input
terminal on the rear panel. [POS] means positive polarity, and triggers are
generated by detecting the rise of the TTL signal. [NEG] means negative
polarity, and triggers are generated by detecting the fall of the TTL signal.
390
[Trigger]
Function
Function
key
GPIB command
HOLD
SING
NUMG <value>
CONT
(N/A)
GPIB command
TRGS INT
TRGS EXT
TRGS BUS
TRGS MAN
TRGEVE {SWE|POIN}
TRGP {POS|MEG}
Appendix B