Preparation of Measurement Accessories
Adapter Setting
NOTE
When the load data measurement is completed, the softkey label changes to
OPEN [DONE]
Step 8. Connect the SHORT (0 W, HRM 504) to the 3.5-mm port (the SHORT is furnished with the
42941A Impedance Probe).
Step 9. Press the
NOTE
When the short data measurement is completed, the softkey label changes to
SHORT [DONE]
Step 10. Remove the SHORT (0 W, HRM 504) from the 3.5-mm port of the 42941A. Then connect
the LOAD (50 W, HRM 601) to the 3.5-mm port (the LOAD is furnished with the 42941A
Impedance Probe).
Step 11. Press the
NOTE
When the load data measurement is completed, the softkey label changes to
LOAD [DONE]
Step 12. Press the
compensation, OPEN, SHORT, and LOAD data. The adapter setup data is automatically
saved to the EEPROM (nonvolatile memory) in the Agilent 4294A.
NOTE
Press the
canceled, the previously obtained data is saved as the available adapter setup data.
80
.
key to start the short data measurement.
SHORT [-]
.
key to start the load data measurement.
LOAD [-]
.
key to start calculating the adapter setup data from the measured phase
done
key to cancel the adapter setup process. If the adapter setup process is
cancel
Chapter 4