Calibration
G. Calibration When the 3.5-mm Port of the 42941A is Extended
G. Calibration When the 3.5-mm Port of the 42941A is
Extended
Perform the following calibration when the 3.5-mm port of a 42941A connected to the
Agilent 4294A is extended with a one-port configuration and the device is connected to the
cable's end with a minimal connection.
Figure 6-7

Calibration When the 3.5-mm Port of the 42941A is Extended

Step 1. Make sure that the adapter selection is set to "42941A Impedance Probe (PROBE
42941A)."
See "Adapter Setting" on page 67 for adapter selection.