Calibration
Fixture Compensation
Step 4. Select one of the compensation data definition keys,
SHORT RESIST(R)
confirm or set the compensation data definition.
Compensation Data Definition
Equivalent parallel conductance value (G) of
OPEN standard used for fixture compensation
(preset state: 0 S)
Equivalent parallel capacitance value (C) of
OPEN standard used for fixture compensation
(preset state: 0 F)
Equivalent series resistance value (R) of
SHORT standard used for fixture
compensation
(preset state: 0 W)
Equivalent series inductance value (L) of
SHORT standard used for fixture
compensation
(preset state: 0 H)
Equivalent series resistance value (R) of
LOAD standard used for fixture compensation
(preset state: 50 W)
Equivalent series inductance value (L) of
LOAD standard used for fixture compensation
(preset state: 0 H)
The current value of the compensation data definition is displayed in the setting parameter
area of the LCD's upper-left side.
Step 5. Use the keys or rotary knob of the ENTRY block in one of the following ways to specify
the compensation data definition .
NOTE
When entering the capacitance value in the compensation data definition, the
automatically added at the end of the value you enter with the unit keys. For example, if
you enter a value with the numeric keys and then press the
the entered value.
Step 6. Repeat steps 4 and 5 to set all of the required compensation data definitions.
154
,
SHORT INDUCT(L)
Enter the desired value with the numerical keys (
one of the unit keys (
[G/n]
Turn the rotary knob (
) until the desired value is set.
m m m m
Press the step keys (
) to set the desired value.
[­][¯]
OPEN CONDUCT(G)
,
LOAD RESIST(R)
Key Stroke
OPEN CONDUCT(G)
OPEN CAP(C)
SHORT RESIST(R)
SHORT INDUCT(L)
LOAD RESIST(R)
LOAD INDUCT(L)
[0]
,
,
, or
).
[M/m]
[k/m]
[´1]
,
OPEN CAP(C)
, or
LOAD INDUCT(L)
through
,
,
) and then press
[9]
[.]
[-]
unit is
[fF]
key, the
unit is added to
[x1]
[fF]
Chapter 6
,