Agilent Technologies 4294A Operating Manual page 409

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

Table B-69
Search range (Search Rng) menu (1)
Key operation:
[Search] - SEARCH RANGE MENU
Softkey
MKR ® MAXIMUM
Substitutes the sweep parameter value at the main marker on the current active
trace (A or B) into the partial search range upper limit. In other words, the sweep
range is changed using the current position of the main marker as the new partial
search range upper limit.
MKRD ® SEARCH
Uses the range between the main marker on the current active trace (A or B) and
the D marker as the partial search range.
RNG
return
Returns to Table B-64, "Search (Search) menu," on page 397.
1. When the sweep type is not list (list sweep).
Table B-70
Search range (Search Rng) menu (2)
Key operation:
[Search] - SEARCH RANGE MENU
Softkey
FULL
Selects the mode in which the partial search range is the entire list sweep table.
|
SEGMENT
Selects the mode in which the partial search range is one specific segment in the
list sweep table.
SEGMENT NUMBER
When a segment is selected as the partial search range, specify a segment
number for partial search.
RNG LINE on OFF
Toggles on and off the partial search range border line display to indicate the
borders of the partial search range by two vertical lines on the graph. If the
measurement parameter is COMPLEX Z-Y, you cannot turn on the partial search
range border line display.
return
Returns to Table B-64, "Search (Search) menu," on page 397.
1. When the sweep type is list (list sweep).
Appendix B
Softkeys displayed by pressing the [Search] key
1
Function
1
Function
Key Definitions
GPIB command
MKRTRMAX
MKRTR
(N/A)
GPIB command
SEARNG FULL
SEARNG SEGMENT
SEGMNUM <value>
SEARLINE {ON|OFF}
(N/A)
401