Agilent Technologies 4294A Operating Manual page 175

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

Figure 7-7
Positioning the reference line
NOTE
The reference line is a horizontal dotted line displayed in the same color as the
corresponding trace. At the left end of the reference line, a " "symbol is displayed in the
same color.
Step 5. Press the
the screen indicates the reference value currently in effect, that is, the value of the
measurement parameter that corresponds to the current reference line position.
Step 6. Use the keys or rotary knob of the ENTRY block in one of the following ways to specify
the reference value.
Enter the desired value with the numeric keys (
the unit keys (
Turn the rotary knob (
Press the step keys ([­][¯]) to set the desired value.
Chapter 7
Setting Up the Display of Measurement Results
Manual Scale Setting (for measurements other than COMPLEX Z-Y)
key. The Parameter Setting field in the upper left area of
REFERENCE VALUE
,
,
, or
[G/n]
[M/m]
[k/m]
m m m m
) until the desired value is set.
to
and
) and then press one of
[0]
[9]
[.]
).
[´1]
167