Agilent Technologies 4294A Operating Manual page 157

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

Step 2. Press the
Step 3. Press the
Step 4. Select one of the calibration data definition keys,
SHORT RESIST(R)
confirm or set the calibration data definition.
Calibration Data Definition
Equivalent parallel conductance value (G) of
OPEN standard used for user calibration
(preset state: 0 S)
Equivalent parallel capacitance value (C) of
OPEN standard used for user calibration
(preset state: 0 F)
Equivalent series resistance value (R) of
SHORT standard used for user calibration
(preset state: 0 W)
Equivalent series inductance value (L) of
SHORT standard used for user calibration
(preset state: 0 H)
Equivalent series resistance value (R) of
LOAD standard used for user calibration
(preset state: 50 W)
Equivalent series inductance value (L) of
LOAD standard used for user calibration
(preset state: 0 H)
The current value of the calibration data definition is displayed in the parameter setting
area of the LCD's upper-left side.
Step 5. Use the keys or rotary knob of the ENTRY block in one of the following ways to specify
the calibration data definition.
Enter the desired value with the numerical keys (
one of the unit keys (
Turn the rotary knob (
Press the step keys (
NOTE
When the capacitance value of the calibration data definition is entered with the numerical
keys and one of the unit keys, the unit [fF] is automatically added to the entered value.
Step 6. Repeat procedures 4 and 5 to set all of the required calibration data definitions.
Chapter 6
key to display the User Calibration Menu.
USER CAL
key to display the Calibration Data Definition Menu.
DEFINE VALUE
,
SHORT INDUCT(L)
,
,
[G/n]
[M/m]
[k/m]
) until the desired value is set.
m m m m
) to set the desired value.
[­][¯]
OPEN CONDUCT(G)
,
, or
LOAD RESIST(R)
LOAD INDUCT(L)
Key Stroke
OPEN CONDUCT(G)
OPEN CAP(C)
SHORT RESIST(R)
SHORT INDUCT(L)
LOAD RESIST(R)
LOAD INDUCT(L)
through
,
[0]
[9]
[.]
, or
).
[´1]
Calibration
User Calibration
,
,
OPEN CAP(C)
, to
,
) and then press
[-]
149