2. Oscillator Calibration (OSC CAL)
Required Equipment
Power Meter
Power Sensor
N(f)-BNC(m) Adapter
Procedure
1. Connect the equipment as shown in Figure 2-4.
2. Select \OSC CAL" and click
3. Following the program instructions, complete the procedure.
2-10 Adjustments and Performance Tests
436A, 437B, or 438A
8482A
PN 1250-1477
Figure 2-4. Oscillator Calibration Setup
NNNNNNNN
OK in the list box of the program display.