Table B-83
Limit line edit (Edit Limit) menu
Key operation:
[System] - LIMIT TEST - EDIT LIMIT LINE
Softkey
SEGMENT
From the limit line table, specifies a segment number to be edited. The currently
specified segment is indicated by the pointer (>) on the table. You can change the
specified segment number using the ENTRY block keys (numeric key, rotary
knob, [­][¯] keys) on the front panel. If there is no segment in the limit line table
("EMPTY" is displayed at the top of the table), add/edit segments using the
EDIT key or ADD key.
EDIT ä
For the segment indicated by the current pointer (enters into the edit mode. In the
segment edit mode, the menu changes to Table B-84, "Segment edit (Edit Limit)
menu (1/3)," on page 409. If there is no segment in the limit line table
("EMPTY" is displayed at the top of the table), a preset segment is added as
substitution and the edit mode is invoked.
DELETE
Deletes the segment indicated by the current pointer (>).
ADD ä
Inserts a segment whose contents are the same as the segment indicated by the
current pointer (>) after the segment and enters the segment edit mode. In the
segment edit mode, the menu changes to Table B-84, "Segment edit (Edit Limit)
menu (1/3)," on page 409. If there is no segment in the limit line table
("EMPTY" is displayed at the top of the table), a preset segment is added as
substitution and the edit mode is invoked.
CLEAR LIST ä
Displays Table B-87, "List deletion (Clear List?) menu," on page 412.
done
Finishes the editing of the limit line table, replaces the table before the editing
with the new table, and stores it. Also, returns to Table B-82, "Limit test (Limit
Test) menu," on page 408.
cancel
Cancels the editing of the limit line table and returns to Table B-82, "Limit test
(Limit Test) menu," on page 408. The limit line table before the editing remains
unchanged.
Table B-84
Segment edit (Edit Limit) menu (1/3)
Key operation:
[System] - LIMIT TEST - EDIT LIMIT LINE - EDIT
[System] - LIMIT TEST - EDIT LIMIT LINE - ADD
Softkey
START
Sets the sweep start value of the segment.
UPPER LIMIT
Sets the upper limit of the measured value at the sweep start point of the
segment
limit is not required, set an extremely large value (for example, 1G) as the upper
limit. If you set the upper limit smaller than the lower limit, both limits are set to
the newly entered value automatically.
LOWER LIMIT
Sets the lower limit of the measured value at the sweep start point of the
segment
is not required, set an extremely small value (for example, -1G) as the lower
limit. If you set the lower limit greater than the upper limit, both limits are set to
the newly entered value automatically.
Appendix B
Softkeys displayed by pressing the [System] key
Function
Function
1
. Be sure to set both the upper limit and the lower limit. If an upper
1
. Be sure to set both the upper limit and the lower limit. If a lower limit
Key Definitions
GPIB command
LIMSEGM <value>
LIMSEDI
LIMSDEL
LIMSADD
LIMCLEL
LIMEDONE
(N/A)
GPIB command
LIMSTAR <value>
LIMUSTAR <value>
LIMLSTAR <value>
409