Setting Measurement Conditions
Using Time as Sweep Parameter (Zero Span Sweep)
-
-
keys.
[Utility]
MKR X-AXIS [ ]
TIME
Sweep parameters other than frequency can be selected in Step 3. In such a case, the fixed
value for the signal source is reflected in the selected parameter in Steps 5 and 6.
Settings for signal sources other than the fixed value in Steps 5 and 6 can be made in the
signal source menu, which can be accessed by pressing the
key.
[Source]
The zero span sweep can also be achieved with the
and
keys to specify the
[Start]
[Stop]
sweep start value and sweep stop value as the same value.
Figure 5-10
Sweep span value = 20 kHz
Chapter 5
95